It was reported that a revision procedure to extend an existing construct was performed on an unknown date, utilizing the reform pedicle screw system.While removing a previously implanted screw, the tip of the t20, polyaxial driver, reform degen (39-sp-0700) broke.The tip was removed and the procedure was competed utilizing other instrumentation readily available in the set.There was no patient injury or delay to the procedure.
|
H3 device evaluation - the driver is fractured at the bottom of the hexalobe.The fracture plane is normal to the driver's longitudinal axis.This type of fracture is consistent with a failure due to torsional overload.It is unclear if the driver was fully threaded / tightened to the screw or if the driver tip was just inserted when torque was applied.Fully tightening would enable some load sharing.The part has likely seen extensive use since being fabricated in 2015 so this failure may have been initiated during prior usage.Review of device history records (ir15-1284) found thirty (30) pieces of lot 33611mm were received from the supplier, medical machining specialists, and released for distribution on 9/16/2015 with no deviation or anomalies.Two-year complaint history review did not reveal a trend for reports of this nature for this part number.No corrective actions are recommended.
|