RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60 V60PLUS VENTILATOR
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/20/2024
Event Type  malfunction  
Event Description
Philips received a complaint from the customer, reporting that the v60 ventilator displayed error code 110a ((post) check vent: proximal pressure sensor auto zero failed).There was no patient involvement when the issue occurred.No patient or user harm reported.A philips remote service engineer (rse) noted that the customer's v60 ventilator displayed error code 110a ((post) check vent: proximal pressure sensor auto zero failed).During troubleshooting, the rse noted that the customer confirmed the error code in device event log; however, the customer was unable to recreate error.It was then noted that the customer ran the device overnight, and device did not have any error codes.The rse provided the customer with the following instructions per the manufacturer's recommendation, verify pin alignment between solenoids and the data acquisition (da) board, replace the proximal auto-zero solenoid (sol 3 and sol 4), replace the da board.The customer was provided with the part number for a replacement solenoid valves.
 
Manufacturer Narrative
H10: a follow up was performed with the customer, and it was reported that the solenoid valves (3 and 4) were replaced to resolve the issue.The device was tested and operates as intended.The device was returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18998146
MDR Text Key338914215
Report Number2518422-2024-15915
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838090996
UDI-Public(01)00884838090996
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1138747
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/28/2024
Supplement Dates Manufacturer Received04/04/2024
Supplement Dates FDA Received04/11/2024
Date Device Manufactured11/03/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-