• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PERCUFLEX URINARY DIVERSION STENT; CATHETER, NEPHROSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PERCUFLEX URINARY DIVERSION STENT; CATHETER, NEPHROSTOMY Back to Search Results
Model Number M0061603300
Device Problems Break (1069); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/09/2024
Event Type  malfunction  
Manufacturer Narrative
H6: imdrf device code a0401 captures the reportable investigation results of stent detached/separated.H10: the returned percuflex urinary diversion stent was analyzed, and a visual and microscopic evaluation noted that the coil was kinked and cut at the shaft.No other problems with the device were noted.The reported event was confirmed.Taking all available information into consideration, most likely, the device meets all manufacturing specifications required and passed all the controls and inspections, no abnormalities were reported during the assembly process.It is probable that the problem was caused due to the interaction between the guidewire and the stent device as well, such as the handling of the device could have contributed with the complaint event reported.Regarding founded issue of stent shaft detached, based on the irregular shape of the detached tip, there is enough evidence to conclude that the device was cut.Therefore, all compiled information on this investigation determines that the most probable cause is unintended use error cause or contributed to event since the interaction between the user and device, or sample, caused or contributed to the error.
 
Event Description
It was reported to boston scientific corporation that a percuflex urinary diversion stent was used during a procedure performed on (b)(6) 2024.During preparation, the stent was damaged.It was found that the pigtail part was kinked.The procedure was successfully completed with another percuflex urinary diversion stent.There were no patient complications reported as a result of this event.Investigation results revealed that the stent was detached/separated, therefore this event has been deemed an mdr reportable event.Please see block h10 for full investigation details.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCUFLEX URINARY DIVERSION STENT
Type of Device
CATHETER, NEPHROSTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key19000475
MDR Text Key338901135
Report Number2124215-2024-17768
Device Sequence Number1
Product Code LJE
UDI-Device Identifier08714729821144
UDI-Public08714729821144
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K830803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM0061603300
Device Catalogue Number160-330
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/28/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-