• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER AU5800 CLINICAL CHEMISTRY ANALYZER; ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER AU5800 CLINICAL CHEMISTRY ANALYZER; ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE Back to Search Results
Model Number AU5800-10 CHEMISTRY ANALYZER AU5800
Device Problem Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/05/2024
Event Type  malfunction  
Manufacturer Narrative
The customer performed their own troubleshooting over the phone with beckman technical support and replaced the sample probes to resolve the issue.The customer verified the analyzer resumed to normal operation.Section a2, a4, and a5: information not provided by customer.Beckman coulter internal identifier is: (b)(4).
 
Event Description
The customer reported obtaining negative patient results for total bilirubin and rheumatoid factor latex on their au5800 clinical chemistry analyzer.There was no report of change to treatment, injury, or death associated with this event.The customer did not provide patient data or demographics.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AU5800 CLINICAL CHEMISTRY ANALYZER
Type of Device
ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd.
brea CA 92821 8000
Manufacturer (Section G)
BECKMAN COULTER MISHIMA K.K.
454-32 higashino
nagaizumi-cho sunto-gun
schizuoka JP-NO TA 4
JA   JP-NOTA 41
Manufacturer Contact
angela vettel
250 s. kraemer blvd.
brea, CA 92821-8000
7149613625
MDR Report Key19000970
MDR Text Key338997537
Report Number9612296-2024-00116
Device Sequence Number1
Product Code JJE
UDI-Device Identifier14987666543016
UDI-Public(01)14987666543016(11)210517
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112412
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAU5800-10 CHEMISTRY ANALYZER AU5800
Device Catalogue NumberB96697
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/05/2024
Initial Date FDA Received03/28/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/17/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-