Model Number 96280 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/28/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The spacelabs health care field service engineer (fse) inspected all the relevant network hardware and could not find any faults.Further on site testing is expected with additional assistance from a product support specialist.A follow up report will be submitted in accordance with 21 cfr 803.56 when additional information becomes available.
|
|
Event Description
|
A spacelabs healthcare field service engineer (fse) was on site and stated that they observed several telemetry beds going offline for several minutes before coming back online on its own.There was no patient or user harm associated with this event.
|
|
Manufacturer Narrative
|
The hospital network infrastructure was inspected and it was suspected that the network hardware was the source of the outage.Updates, configuration changes, and network hardware was replaced to improve the customer network.The customer confirmed that the improvements to their network have seemed to resolve the intermittent loss of network communication.
|
|