SYNTHES PRODUKTIONS GMBH 2.0MM TI ELASTIC NAIL 440MM; PIN, FIXATION, SMOOTH Back to Search Results
Catalog Number 475.920
Device Problem Break (1069)
Patient Problem Insufficient Information (4580)
Event Date 01/23/2024
Event Type  malfunction  
Event Description
Device report from china reports an event as follows: it was reported that on (b)(6) 2024 the screw in question broke during surgery.During the plastic forming process, the screw broke.Another device of a similar type was used to complete the procedure.There were no adverse patient consequences.No additional medical intervention was required.No additional information is available.This report is for a 2.0mm ti elastic nail 440mm.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Part: 475.920 lot no: 94p8338 release to warehouse date: 23 mar 2021 manufacturing site: werk selzach supplier: (b)(4).The product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed that the ten ø2 l440 tan green was found broken at the middle section of the shaft.No other issues were observed.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed for ten ø2 l440 tan green.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: d9 h3, h6: a product investigation was conducted.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the ten ø2 l440 tan green was found broken at the middle of the nail.A dimensional inspection was not performed due to the post-manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.The overall complaint was confirmed as the observed condition of the ten ø2 l440 tan green would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Corrected data: d4: corrected udi number.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.0MM TI ELASTIC NAIL 440MM
Type of Device
PIN, FIXATION, SMOOTH
Manufacturer (Section D)
SYNTHES PRODUKTIONS GMBH
dornacherstrasse 20
balsthal 4710
SZ  4710
Manufacturer (Section G)
WERK SELZACH
bohnackerweg 5
selzach 2545
SZ   2545
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key19003416
MDR Text Key338990718
Report Number8030965-2024-04233
Device Sequence Number1
Product Code HTY
UDI-Device Identifier07611819093027
UDI-Public(01)07611819093027
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K971783
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number475.920
Device Lot Number94P8338
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/13/2024
Initial Date FDA Received03/29/2024
Supplement Dates Manufacturer Received04/18/2024
Supplement Dates FDA Received05/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/23/2021
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-