H9 correction and removal number: 3002808148-10/09/2023-001-c this supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Additional information added to field: h3, h4, h6, h7, and h9.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 10 years since the subject device was manufactured.Based on the results of the investigation, it is presumed that the event occurred due to detection of error in the pressure sensor on the main printed circuit board (pcb) which stopped the insufflation operation while sounding the alarm tone, and that light- emitting diode (led) other than the power supply turned off.However, a definitive root cause could not be determined.Three attempts were performed to obtain additional information, but no response was received from the customer.Olympus will continue to monitor field performance for this device.
|