• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/13/2024
Event Type  malfunction  
Event Description
It was observed that during the device evaluation, the high flow insufflation unit exhibited an error alert that didn¿t stop when the device was turned on.There were no reports of patient involvement.
 
Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H9 correction and removal number: 3002808148-10/09/2023-001-c this supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Additional information added to field: h3, h4, h6, h7, and h9.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 10 years since the subject device was manufactured.Based on the results of the investigation, it is presumed that the event occurred due to detection of error in the pressure sensor on the main printed circuit board (pcb) which stopped the insufflation operation while sounding the alarm tone, and that light- emitting diode (led) other than the power supply turned off.However, a definitive root cause could not be determined.Three attempts were performed to obtain additional information, but no response was received from the customer.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19003932
MDR Text Key339041060
Report Number3002808148-2024-03002
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Repair
Type of Report Initial,Followup
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/29/2024
Initial Date FDA Received03/29/2024
Supplement Dates Manufacturer Received04/10/2024
Supplement Dates FDA Received04/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/11/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-