Catalog Number 2C8750
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/29/2024
Event Type  malfunction  
Event Description
It was reported that a clearlink system y-type blood/solution set "snapped".This was observed after the platelet infusion completed; the platelet side of blood tubing was clamped and the saline side was opened.The platelet bag fell and the bag was "was still spiked, and the tubing had snapped".There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: two (2) actual devices were received for evaluation.A visual inspection was performed and observed that the spike was separated from the tubing for both samples.The reported condition was verified.The cause of the condition was determined to be related to an assembly issue during manufacturing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK BLOOD RECIPIENT SET
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
carretera sanchez km 18.5
parque industrial itabo, piisa
haina, san cristobal 91000
DR   91000
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key19004111
MDR Text Key338934189
Report Number1416980-2024-01414
Device Sequence Number1
Product Code BRZ
UDI-Device Identifier00085412046341
UDI-Public(01)00085412046341
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K993120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C8750
Device Lot NumberDR23L07024
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 03/04/2024
Initial Date FDA Received03/29/2024
Supplement Dates Manufacturer Received05/02/2024
Supplement Dates FDA Received05/03/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/08/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
PLATELETS; SALINE
-
-