Catalog Number 2C8750 |
Device Problem
Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/29/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that a clearlink system y-type blood/solution set "snapped".This was observed after the platelet infusion completed; the platelet side of blood tubing was clamped and the saline side was opened.The platelet bag fell and the bag was "was still spiked, and the tubing had snapped".There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
H10: two (2) actual devices were received for evaluation.A visual inspection was performed and observed that the spike was separated from the tubing for both samples.The reported condition was verified.The cause of the condition was determined to be related to an assembly issue during manufacturing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|