Model Number DQ8001
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problem Burn(s) (1757)
Event Date 03/08/2023
Event Type  Injury  
Event Description
Severely burned patient unit needs to be checked out by vendor.Pre-existing medical conditions is paraplegia, mild edema.Device is 4 years old and used for 4 years.No other devices were being used at the time of event.No discomfort at time or after incident.Used in ifc 40 ma.Unit last serviced 5/18/2022 prior to incident.They are using 2" by 4" self adhesive electrodes on b anterior tibialis, 2 each le, 4" apart.New electrodes , only one use.Nothing was applied to the skin prior and device was turned off when pads were removed/adjusted.Skin was clear.Burns appeared the next day, (b)(6) 2023, and resolved on (b)(6) 2023.They do not have experience with this waveform.The patient did seek medical attention and they have the medical records.Electrodes are not shared between multiple patients and are stored in original package in therapy gym.They were using ifc, 4 electrodes, 2 per le.Patient has no known allergies.They are using 2" x 4" dura-stick plus electrodes.They have not tried new leads, unit is out of service after incident.Leads show no damage.User manual attached.Page 6, warning: do not use this device under these conditions, 5th bulletin: "· over areas of skin that lack normal sensation." warning (cont.), general warnings:, page 8, 6th bulletin: "before administering any treatment to a patient you should become acquainted with the operating procedures for each mode of treatment available, as well as the indications, contraindications, warnings and precautions.Consult other resources for additional information regarding the application of electrotherapy and ultrasound." page 8, electrotherapy warnings, 2nd bulletin: "this device is to be used as a symptomatic treatment for pain and has no curative value.Patients should be cautioned and their activities regulated if pain that would otherwise serve as a protective mechanism is suppressed.".On 01-31-2024 customer reported "severely burned patient unit needs to be checked out by vendor" with no response until 03/08/2024 where they provided the additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSCOE MEDICAL
Type of Device
THERATOUCH CX4 W/THERAPY CART
Manufacturer (Section D)
SHENZHEN DONGDIXIN TECHNOLOGY CO., LTD.
floor 1-2, no.3 building, fans
industrial estate xilixiaobaim
shenzhen guangdong, 51810 8
CH  518108
MDR Report Key19005934
MDR Text Key338981898
Report Number3012316249-2024-00006
Device Sequence Number1
Product Code GYB
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberDQ8001
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/29/2024
Distributor Facility Aware Date03/08/2024
Device Age56 MO
Event Location Other
Date Report to Manufacturer03/29/2024
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/29/2024
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age51 YR
Patient Weight78 KG
-
-