• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-H180J
Device Problem Device Reprocessing Problem (1091)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2024
Event Type  malfunction  
Event Description
It was reported the customer failed to inspect the leakage tester and maintenance unit before connecting to the gastrointestinal videoscope.The issue occurred during reprocessing.There were no reports of patient harm.
 
Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over twelve (12) years since the subject device was manufactured.An olympus endoscopy support specialist (ess) was dispatched on 11 march 2024, to observe the user facility¿s reprocessing practices from start to finish and provide a reprocessing in-service training, if necessary, to correct and address any reprocessing deviations.During the in-service the ess covered the guidelines on reprocessing the olympus scope per reprocessing manual.The staff also performed a return demonstration to show they understood the process.The customer also understood that the olympus reprocessing manuals are the validated source of instructions.A definitive root cause was not identified, however based on the results of the investigation, the probable cause of the "when the user performed leakage test, they did not check mb-155-leakage tester or mu-1-maintenance unit prior to connection to the scope" malfunctions would likely be related to difference in recognition on device handling or reprocessing steps between olympus recommendation and the user facility.The event can be prevented by following the instructions for use which state: evis exera ii gif/cf/pcf type 180 series reprocessing manual chapter 5 reprocessing the endoscope (and related reprocessing accessories).Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19006045
MDR Text Key339335571
Report Number9610595-2024-06690
Device Sequence Number1
Product Code FDS
UDI-Device Identifier04953170275456
UDI-Public04953170275456
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051645
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIF-H180J
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/11/2024
Initial Date FDA Received03/29/2024
Supplement Dates Manufacturer Received04/11/2024
Supplement Dates FDA Received04/18/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/13/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-