The patient was undergoing a thrombectomy procedure in the peripheral arterial vasculature using an indigo system catrx aspiration catheter (catrx) and a non-penumbra guide catheter.During the procedure, the physician completed few passes using the catrx.While treating a right side peripheral vessel, the catrx fractured at the proximal end.The catrx was removed intact with the guide catheter using forceps.It is unknown how the procedure was completed.There was no report of an adverse effect to the patient.
|
Evaluation of the returned catrx confirmed a fracture on its proximal shaft.If the catrx is retracted at an angle during use, damage such as a kink and subsequent fracture may occur.Further evaluation of the device revealed additional kinks, bends along the length of the catheter shaft, and multiple kinks near the distal tip of the guidewire lumen.This damage was incidental to the complaint and likely occurred during removal from patient¿s body and packaging for return to penumbra.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|