This report is being submitted retrospectively as part of internal review.Transmitter was not requested to be returned for evaluation as it was still being used by the user.Investigation was performed on user's data available in data management system (dms).Based on the investigation analysis, the reported event (the fingerstick measurement of 57 mg/dl and the corresponding sensor reading of 91 mg/dl) could not be confirmed in dms which was reported around 1:04 pm cet on (b)(6) 2022.At around 1:01 pm cet on (b)(6) 2022, the system displayed a sensor reading of 126 mg/dl and a calibration entry of 120 mg/dl.This shows the system was in good agreement with the fingerstick measurements.The overall sensor performance was within expectations.No further investigation was necessary.
|