MAUDE Adverse Event Report: SENSEONICS INC. EVERSENSE SMART TRANSMITTER; IMPLANTABLE GLUCOSE MONITORING SYSTEM, TRANSMITTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SENSEONICS INC. EVERSENSE SMART TRANSMITTER; IMPLANTABLE GLUCOSE MONITORING SYSTEM, TRANSMITTER Back to Search Results
Model Number 102208-704
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2022
Event Type  malfunction  
Event Description
Senseonics was made aware of an incident where patient complained that transmitter got extremely hot while charging and would not turn on.Incident did not lead to any injuries.
 
Manufacturer Narrative
This report is being submitted retrospectively as part of internal review.Transmitter was requested to be returned for evaluation.The transmitter was initially tested for visual inspection which did not reveal any anomalies.The unit was then tested for functional verification and it passed all the tests.As a result, the customer complaint could not be confirmed or the problem could not be reproduced in lab testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SMART TRANSMITTER
Type of Device
IMPLANTABLE GLUCOSE MONITORING SYSTEM, TRANSMITTER
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20876-7005
MDR Report Key19009360
MDR Text Key339010186
Report Number3009862700-2024-00319
Device Sequence Number1
Product Code QCD
Combination Product (y/n)Y
Reporter Country CodeGM
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 03/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/06/2021
Device Model Number102208-704
Device Catalogue NumberFG-3400-04-001
Device Lot Number121136
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2022
Initial Date Manufacturer Received 01/18/2022
Initial Date FDA Received03/31/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/06/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-