Catalog Number 08791732190 |
Device Problems
High Test Results (2457); Non Reproducible Results (4029)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/16/2024 |
Event Type
malfunction
|
Event Description
|
The initial reporter stated they received discrepant results for one patient sample tested with elecsys total psa on a cobas e 801 analytical unit.The sample initially resulted in a total psa value of 2.14 ng/ml.The sample was repeated, resulting in a total psa value of < 0.00600.The repeat value was clinically expected for the patient.
|
|
Manufacturer Narrative
|
The serial number of the e801 analyzer is (b)(6).The investigation is ongoing.
|
|
Manufacturer Narrative
|
Calibration and quality controls were acceptable.No relevant alarms were identified on the alarm trace data.Based on the data provided a general reagent issue was excluded.The reagent performed within specifications.No product issue was found.The root cause of the event could not be identified.
|
|