ROCHE DIAGNOSTICS ELECSYS TOTAL PSA; PROSTATE-SPECIFIC ANTIGEN (PSA) FOR MANAGEMENT OF PROSTATE CANCERS Back to Search Results
Catalog Number 08791732190
Device Problems High Test Results (2457); Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/16/2024
Event Type  malfunction  
Event Description
The initial reporter stated they received discrepant results for one patient sample tested with elecsys total psa on a cobas e 801 analytical unit.The sample initially resulted in a total psa value of 2.14 ng/ml.The sample was repeated, resulting in a total psa value of < 0.00600.The repeat value was clinically expected for the patient.
 
Manufacturer Narrative
The serial number of the e801 analyzer is (b)(6).The investigation is ongoing.
 
Manufacturer Narrative
Calibration and quality controls were acceptable.No relevant alarms were identified on the alarm trace data.Based on the data provided a general reagent issue was excluded.The reagent performed within specifications.No product issue was found.The root cause of the event could not be identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS TOTAL PSA
Type of Device
PROSTATE-SPECIFIC ANTIGEN (PSA) FOR MANAGEMENT OF PROSTATE CANCERS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key19014133
MDR Text Key339046843
Report Number1823260-2024-00986
Device Sequence Number1
Product Code LTJ
UDI-Device Identifier07613336165980
UDI-Public07613336165980
Combination Product (y/n)Y
Reporter Country CodeIT
PMA/PMN Number
P990056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number08791732190
Device Lot Number75666301
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/16/2024
Initial Date FDA Received04/01/2024
Supplement Dates Manufacturer Received05/06/2024
Supplement Dates FDA Received05/08/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-