As reported, the white section of a 6f/7f mynxcontrol vascular closure device (vcd) split when the tip of the device was inserted, and the device was not able to be inserted into the unknown sheath.A new mynx device was opened and the vessel was closed without issue.There was no reported patient injury.The device was properly prepped and inspected prior to use.The device will be returned for evaluation.Addendum: product analysis demonstrates that the sealant was found slightly exposed from the sealant sleeves and exposed to blood.
|
As reported, the white section of a 6f/7f mynx control vascular closure device (vcd) split when the tip of the device was inserted, and the device was not able to be inserted into the unknown sheath.A new mynx device was opened and the vessel was closed without issue.There was no reported patient injury.The device was properly prepped and inspected prior to use.A non-sterile ¿mynx control vcd 6f-7f¿ involved in the reported complaint was returned for investigation.Visual inspection of the received device showed that both button 1 and button 2 were not depressed.The procedure sheath was not returned for evaluation and the syringe was received connected to the device with the stopcock observed opened.In addition, the balloon was found fully deflated.The sealant was found slightly exposed from the sealant sleeves and exposed to blood.The sealant sleeves were observed to have been kinked/bent; however, no cracks were observed on it.The device was inspected for other damages/anomalies that may have contributed to the reported failure and no damages/anomalies were observed on the returned device.A dimensional test was not performed on the returned device due to the kinked condition observed of the sealant outer sleeve assembly.Per functional analysis, a simulated deployment test was performed on the returned device per the mynx control instructions for use (ifu), step 2: deploy sealant.Button 1 was able to be depressed to deploy the sealant with no resistance felt.No issues were noted with respect to button 1 deployment during the device failure investigation.The returned device performed as intended per the mynx control ifu.Button #2 was able to be fully depressed, and no issues were noted with respect to button 2.Per microscopic analysis, visual inspection at high magnification revealed that the sealant was found slightly exposed from the sealant sleeves and exposed to blood.The sealant sleeves were observed to have been kinked/bent, however, no cracks were observed on it.The reported event of ¿sealant sleeves (cartridge assembly)-frayed/split/torn¿ was not confirmed through analysis of the returned device; however, a kinked/bent condition of the sleeves was noted.Additionally, a condition was noted in the returned device of ¿mynx control system-deployment difficulty-premature¿ due to the exposed sealant from the kinked sleeves.The exact cause of the observed conditions could not be conclusively determined during analysis.Based on the limited information available for review and product analysis, it is difficult to determine what factors may have contributed to the issue experienced.However, procedural/handling factors (such as using excessive force during insertion into the sheath and/or using an incorrect insertion angle), and/or the condition of the sheath (although not returned) possibly contributed to the damaged condition of the sealant sleeves, the impedance experienced during insertion, and the subsequent premature exposure of the sealant.It should be noted that the mynx control device is manufactured with a slit at the end of the catheter cartridge tubing.The outer sleeve assembly is assembled with 2 side slit overlapping outer sleeves.The sealant is placed right under the outer sleeve assembly and is protected from exposing prematurely.The slits on the outer sleeve assembly are designed to decrease unsheathing force and increase deployment reliability.Refer to the diagram of the mynx control vcd within the ifu displaying the sealant sleeve with slit.If the outer sleeve is damaged/shredded during prepping phase and/or insertion into sheath, it could cause the sealant to be exposed/swollen prematurely and/or obstruct the device path and prevent the device from being inserted into the procedural sheath.As warned in the ifu, which is not intended as a mitigation, ¿do not use if components or packaging appear to be damaged or defective or if any portion of the packaging has been previously opened.¿ additionally, the ifu states ¿step 1: position balloon, insert the mynx control vcd into the procedural sheath through the sheath valve.Advance the catheter until the sheath catch nears the hub of the sheath.Rotate the sheath catch as needed to hook onto the side port of the procedural sheath.¿ neither the product analysis, nor the information available for review suggest that the failures could be related to the design or manufacturing process of the unit.Therefore, no corrective/preventative actions will be taken at this time.
|