RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
No Display/Image (1183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/11/2024 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint from the customer, reporting that the v60 ventilator had a dark display.There was no patient involvement when the issue occurred.No patient or user harm reported.A philips remote service engineer (rse) noted that the customer's v60 ventilator had a dark display.It was noted that the customer was in the process of upgrading the display to the second generation; however, the customer reported that the cable that connects the user interface (ui) board to the switch board does not fit.The customer was informed that they previous cable could be used after the display has been upgraded.Investigation is ongoing.
|
|
Manufacturer Narrative
|
H10: a philips remote service engineer (rse) noted that the customer's v60 ventilator had a dark display.It was noted that the customer was in the process of upgrading the display to the second generation; however, the customer reported that the cable that connects the user interface (ui) board to the switch board does not fit.The customer was informed that they previous cable could be used after the display has been upgraded.A follow up was performed with the customer, and it was reported that the cable was not fully connected to the new screen.After connecting the cable with the screen, the problem was fixed.
|
|