Model Number V60
Device Problem No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2024
Event Type  malfunction  
Event Description
Philips received a complaint from the customer, reporting that the v60 ventilator had a dark display.There was no patient involvement when the issue occurred.No patient or user harm reported.A philips remote service engineer (rse) noted that the customer's v60 ventilator had a dark display.It was noted that the customer was in the process of upgrading the display to the second generation; however, the customer reported that the cable that connects the user interface (ui) board to the switch board does not fit.The customer was informed that they previous cable could be used after the display has been upgraded.Investigation is ongoing.
 
Manufacturer Narrative
H10: a philips remote service engineer (rse) noted that the customer's v60 ventilator had a dark display.It was noted that the customer was in the process of upgrading the display to the second generation; however, the customer reported that the cable that connects the user interface (ui) board to the switch board does not fit.The customer was informed that they previous cable could be used after the display has been upgraded.A follow up was performed with the customer, and it was reported that the cable was not fully connected to the new screen.After connecting the cable with the screen, the problem was fixed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key19015068
MDR Text Key339062025
Report Number2518422-2024-16598
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/11/2024
Initial Date FDA Received04/01/2024
Supplement Dates Manufacturer Received04/04/2024
Supplement Dates FDA Received04/18/2024
Date Device Manufactured07/15/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-