RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Failure to Deliver (2338)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/11/2024 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint from the customer on the v60 indicating that the one-way valve for oxygen flow was not functioning properly.It was stated that when the customer turned on the o2 tank and then unplugged the o2 pressure line from the line, they would get pressure oxygen flow through the wall connector and that the one-way valve was not functioning properly.A field service engineer (fse) went onsite and checked the back flow on the y block for worn out parts and tested the unit several times, but the reported issue could not be reproduced.The investigation concludes that no further action is required currently.It is currently unknown if the unit was in patient use at the time of the reported issue.No patient or user harm was reported.Additional information is being requested regarding use.If/ when additional information becomes available, a follow up submission will be submitted.
|
|
Manufacturer Narrative
|
Based on new information received: issue was determined to be identified during set up which is outside of clinical use on a patient.
|
|