MAUDE Adverse Event Report: CARDINAL HEALTH ENTFLX 12FR 43W OSTYL W Y PRT; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH ENTFLX 12FR 43W OSTYL W Y PRT; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 8884721237E
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Foreign Body In Patient (2687)
Event Date 03/14/2024
Event Type  Injury  
Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event Description
The customer reported that the tip of feeding tube was left inside the patient when removed.Per additional information received, the tip was in the oropharynx.After imaging to confirm the presence of the tip, conscious sedation was administered to have it removed using surgical forceps.Injury was limited to the retention of the foreign object which was treated with surgical forceps.No further medical intervention was required.
 
Manufacturer Narrative
Samples were not received for the investigation.The device history record was reviewed and indicated that the product was released accomplishing all quality standards.Because a sample was not returned, we were unable to perform a follow up investigation to include functional and visual evaluations to confirm the reported issue and determine a root cause.However, a corrective and preventive action has been initiated to address the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTFLX 12FR 43W OSTYL W Y PRT
Type of Device
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key19015181
MDR Text Key339058556
Report Number9612030-2024-00053
Device Sequence Number1
Product Code KNT
UDI-Device Identifier10884521583023
UDI-Public10884521583023
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 05/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model Number8884721237E
Device Catalogue Number8884721237E
Device Lot Number2306500264
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/15/2024
Initial Date FDA Received04/01/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/28/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-