ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PROCLAIM¿ 5 ELITE IMPLANTABLE PULSE GENERATOR; SCS IPG Back to Search Results
Model Number 3660
Device Problems Low impedance (2285); Moisture or Humidity Problem (2986)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/19/2024
Event Type  Injury  
Event Description
It was reported that the patient was unable to set their ipg to mri mode due to low impedances.As a result, the patient underwent surgical intervention on (b)(6) 2024 to address the issue.During the procedure when the lead was disconnected, fluid/tissue was seen in the ipg header.The distal tip of the lead was wiped and the header was suctioned, and the lead was reconnected.Nothing was explanted and all issues were resolved post-operatively.
 
Manufacturer Narrative
Date of event is estimated.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM¿ 5 ELITE IMPLANTABLE PULSE GENERATOR
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
heidi syndergaard
6901 preston road
plano, TX 75024
9723098000
MDR Report Key19015976
MDR Text Key339065048
Report Number3006705815-2024-02397
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067031419
UDI-Public(01)05415067031419(10)A000113152(17)230727
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/27/2023
Device Model Number3660
Device Lot NumberA000113152
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/22/2024
Initial Date FDA Received04/01/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/27/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
Patient Weight91 KG
-
-