A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation was informed that during the initial pass of the aquablation procedure for this case, an "e22 - motorpack error" message was encountered midway through treatment.This error was successfully resolved by restarting the display.However, the treatment process was interrupted multiple times by recurring e22 error messages.Subsequently, an "e50-console error" message was also encountered, leading to the discontinuation of the procedure after the first pass.As a result, the treating surgeon transitioned to performing a turp procedure.There were no adverse health consequences to the patient due to this event.
|
The aquabeam handpiece was returned for investigation.A replacement aquabeam handpiece was provided to the account as part of warranty replacement.Functional testing was unable to replicate the reported e22 and e50 errors.The aquabeam handpiece was deconstructed and viewed under magnification.No signs of fluid ingress were observed on the sensor board and the encoder wheel.The log file review was able to confirm the reported failure mode.However, the associated aquabeam motorpack was concluded as the cause of the reported event.The root cause of the reported event is corrosion caused by fluid ingress on the associated aquabeam motorpack, which triggered an error when connected with the aquabeam handpiece.A review of the device history record (dhr) ab2000-b / serial number (b)(6) and aquabeam handpiece / lot number 24c00049 was conducted, which confirmed there was one non-conformance issued to this lot during the manufacturing process that could potentially be related to the reported event.The lot was segregated and affected units were reworked and re-inspected as part of our handpiece final inspection process.Upon re-inspection, the lot met all required specifications and was deemed acceptable to be released for distribution.The aquabeam robotic system user manual, sj-um0101 rev.B, states the following: table 5: system detected errors and faults e22 - motorpack error release foot pedal and click x.1) if error persists, reconnect handpiece to motorpack.2) if error continues, replace handpiece.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|