• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM Back to Search Results
Catalog Number AB2000
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/27/2024
Event Type  malfunction  
Manufacturer Narrative
Root cause of reported event has not yet been established.Investigation by manufacturer is currently in-process.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Event Description
A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation was informed that during the initial pass of the aquablation procedure for this case, an "e22 - motorpack error" message was encountered midway through treatment.This error was successfully resolved by restarting the display.However, the treatment process was interrupted multiple times by recurring e22 error messages.Subsequently, an "e50-console error" message was also encountered, leading to the discontinuation of the procedure after the first pass.As a result, the treating surgeon transitioned to performing a turp procedure.There were no adverse health consequences to the patient due to this event.
 
Manufacturer Narrative
The aquabeam handpiece was returned for investigation.A replacement aquabeam handpiece was provided to the account as part of warranty replacement.Functional testing was unable to replicate the reported e22 and e50 errors.The aquabeam handpiece was deconstructed and viewed under magnification.No signs of fluid ingress were observed on the sensor board and the encoder wheel.The log file review was able to confirm the reported failure mode.However, the associated aquabeam motorpack was concluded as the cause of the reported event.The root cause of the reported event is corrosion caused by fluid ingress on the associated aquabeam motorpack, which triggered an error when connected with the aquabeam handpiece.A review of the device history record (dhr) ab2000-b / serial number (b)(6) and aquabeam handpiece / lot number 24c00049 was conducted, which confirmed there was one non-conformance issued to this lot during the manufacturing process that could potentially be related to the reported event.The lot was segregated and affected units were reworked and re-inspected as part of our handpiece final inspection process.Upon re-inspection, the lot met all required specifications and was deemed acceptable to be released for distribution.The aquabeam robotic system user manual, sj-um0101 rev.B, states the following: table 5: system detected errors and faults e22 - motorpack error release foot pedal and click x.1) if error persists, reconnect handpiece to motorpack.2) if error continues, replace handpiece.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM ROBOTIC SYSTEM
Type of Device
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS CORPORATION
150 baytech dr
san jose CA 95134
Manufacturer (Section G)
PROCEPT BIOROBOTICS CORPORATION
150 baytech dr
san jose 95134
Manufacturer Contact
doria esquivel
150 baytech dr
san jose 95134
MDR Report Key19016401
MDR Text Key339234013
Report Number3012977056-2024-00098
Device Sequence Number1
Product Code PZP
UDI-Device IdentifierB614AB20001
UDI-Public+B614AB20001/16D202210037
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAB2000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/28/2024
Initial Date FDA Received04/01/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/03/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-