Reportable based on device analysis completed on 28mar2024.It was reported that air leakage was suspected.The 90% stenosed target lesion was located in a diffused plaque near the middle section of the right coronary artery.A 10mmx2.25mm wolverine coronary cutting balloon was selected for use.During the procedure, when the pressure pump was pressed, the pressure did not go up.After the balloon was removed from the outside, the pressure did not go up again, and air leakage was suspected but there was no pinhole noted.The device was slowly removed along the guidewire and procedure was completed with another of the same device.No complications reported and patient was stable.However, device analysis revealed a balloon pinhole.
|