• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 24690
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2024
Event Type  malfunction  
Event Description
It was reported that leak at the tip occurred, and balloon could not hold pressure.The target lesion was located in the internal carotid artery.A 6.0mmx30mmx135cm (4f) sterling balloon catheter was advanced for post dilatation during carotid artery stent placement.The pressure was increased to nominal using a pta balloon indeflator.However, the balloon slowly deflated, and the contrast agent appeared to leak out of the balloon.The pressure was reduced and then it was increased again, the pressurized state could not be maintained, and the contrast agent appeared to leak from the tip of the balloon.The procedure was completed with another of the same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING
Type of Device
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key19021067
MDR Text Key339131595
Report Number2124215-2024-16850
Device Sequence Number1
Product Code DQY
UDI-Device Identifier08714729757276
UDI-Public08714729757276
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K141150
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number24690
Device Catalogue Number24690
Device Lot Number0031814765
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/13/2024
Initial Date FDA Received04/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/14/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age66 YR
Patient SexMale
-
-