Model Number 1845010
Device Problems Overheating of Device (1437); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/07/2024
Event Type  malfunction  
Manufacturer Narrative
H3: product analysis found that the bearings were broken and corroded.As bearings were found broken, it was difficult to insert bur in the attachment, so it was not possible to perform pre repair temperature test.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was reported that pre-op, the indigo attachment was heating and the tool was stuck.There was no patient involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IPC® HANDPIECE ATTACHMENT - INDIGO¿ DRILL
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
glen belmer
6743 southpoint drive north
jacksonville, FL 32216
6122713209
MDR Report Key19022680
MDR Text Key339247802
Report Number1625507-2024-00105
Device Sequence Number1
Product Code HBE
UDI-Device Identifier00763000172862
UDI-Public00763000172862
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K081475
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1845010
Device Catalogue Number1845010
Device Lot Number226368754
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/07/2024
Initial Date FDA Received04/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/17/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-