BOSTON SCIENTIFIC CORPORATION ROTAPRO; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number 3243
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/12/2024
Event Type  malfunction  
Event Description
It was reported that device entanglement occurred.The 75% stenosed target lesion was located in the moderately tortuous and moderately calcified mid left anterior descending artery (lad).A 1.25mm rotapro was selected for use.During procedure, the system was advanced over the wire where the revs were set outside the patient.Dynaglide mode was activated, the brake defeat button was pushed and secured with the wire-clip.When the system was advancing into the catheter it was noted that the end of the wire started to tangle due to the wire spinning rapidly.The wireclip was removed and clipped again, and the brake defeat button was pressed again.The system was removed and the vessel was treated with a balloon and stent.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAPRO
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key19024249
MDR Text Key339207994
Report Number2124215-2024-19920
Device Sequence Number1
Product Code MCX
UDI-Device Identifier08714729893356
UDI-Public08714729893356
Combination Product (y/n)N
Reporter Country CodeSF
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3243
Device Catalogue Number3243
Device Lot Number0030016426
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/13/2024
Initial Date FDA Received04/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-