Additional information was added to d9, h3, h4, h6, and h10: h4: the lot was manufactured between november 9, 2021 ¿ november 10, 2021.H10: the device was received for evaluation.A visual inspection via the naked eye was performed, and the tubing line was broken off at the junction of the distal luer lock.A visual inspection under the microscope was performed, and a jagged mark at the broken aea of the tubing was observed which suggested force may have been applied onto the tubing, causing it to break off.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|