• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Catalog Number 19AGFN-756
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/07/2024
Event Type  malfunction  
Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that on 07 march 2024, a 19mm regent aortic mechanical heart valve ( serial: (b)(6)) was chosen for implantation.After the valve was sutured, a tear was observed on the valve ring.There was no leaflet damage, and no dislodgment of any pieces of the valve.The valve was removed and a replacement 19mm regent valve (serial: (b)(6)) was implanted successfully.The patient remained on bypass throughout the procedure and there was no unusual intervention due to the extended bypass time.There was no patient consequences or clinically significant delay.The patient remained hemodynamically stable throughout the procedure.The patient was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REGENT HEART VALVE WITH FLEX CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key19026205
MDR Text Key339177347
Report Number2135147-2024-01469
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734005852
UDI-Public(01)05414734005852(17)270919(10)C00004261
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number19AGFN-756
Device Lot NumberC00004261
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/08/2024
Initial Date FDA Received04/02/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/20/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-