Complainant alleged that while attempting to treat a 60-year-old female patient, the device was unable to obtain an ecg signal via defib pads.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired, however it was not a result of the reported malfunction.
|
The device was returned to zoll medical corporation for evaluation.The defib pads used during the event were not returned for evaluation.The device was put through extensive testing including bench handling, impedance, and defibrillation cycling using test accessories without duplicating the report.A review of the device log observed the ecg signal was never acquired through pads.The start of the case has two pads off messages with no pads on message.This indicates that the device is recognizing the pads are attached to the patient with an impedance between 300-580 ohms.In order to register a pads on message and obtain an ecg signal, the device must detect a valid impedance between 15-300 ohms.This indicates the pads are not making proper contact to the patient.Poor coupling can occur for multiple reasons, including but not limited to, electrode placement, skin preparation, and patient movement.Electrode labeling states the importance of good placement on the patient and provides instruction for proper electrode application technique.Zoll recommends that patients are cleaned and hair is clipped prior to applying electrode pads to assure good coupling of electrode to skin contact.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|