BOSTON SCIENTIFIC NEUROMODULATION VERCISE GENUS; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
|
Back to Search Results |
|
Model Number DB-1216 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Wound Dehiscence (1154); Bacterial Infection (1735); Unspecified Infection (1930)
|
Event Date 12/13/2023 |
Event Type
Injury
|
Event Description
|
It was reported that the deep brain stimulation (dbs) patient underwent a revision procedure to have the full dbs system explanted due to an infection.The patient was doing well postoperatively.
|
|
Manufacturer Narrative
|
Additional suspect medical device components involved in the event: product family: dbs-linear leads, upn: m365db2202450, model: db-2202-45, serial: (b)(6), batch: 7096269.Product family: dbs-linear leads, upn: m365db2202450, model: db-2202-45, serial: (b)(6), batch: 7094384.
|
|
Manufacturer Narrative
|
Additional suspect medical device components involved in the event: product family: dbs-linear leads, upn: m365db2202450, model: db-2202-45, serial: (b)(6), batch: (b)(6).Product family: dbs-linear leads, upn: m365db2202450, model: db-2202-45, serial: (b)(6), batch: (b)(6).
|
|
Event Description
|
It was reported that the deep brain stimulation (dbs) patient underwent a revision procedure to have the full dbs system explanted due to an infection.The patient was doing well postoperatively.Additional information was received that the infection originated in the suture behind the patients ear and spread to the entire dbs system.The patients suture would not dissolve and upon removing the suture, the patient expressed that the site became hot.Upon further investigation, the entire dbs system was compromised.The physician confirmed that the infection was caused by the suture and open wound at the suture site.The patient was placed on intravenous (iv) antibiotics for one month.A culture was performed and confirmed methicillin-resistant staphylococcus aureus (mssa).
|
|