Model Number DB-1216
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Wound Dehiscence (1154); Bacterial Infection (1735); Unspecified Infection (1930)
Event Date 12/13/2023
Event Type  Injury  
Event Description
It was reported that the deep brain stimulation (dbs) patient underwent a revision procedure to have the full dbs system explanted due to an infection.The patient was doing well postoperatively.
 
Manufacturer Narrative
Additional suspect medical device components involved in the event: product family: dbs-linear leads, upn: m365db2202450, model: db-2202-45, serial: (b)(6), batch: 7096269.Product family: dbs-linear leads, upn: m365db2202450, model: db-2202-45, serial: (b)(6), batch: 7094384.
 
Manufacturer Narrative
Additional suspect medical device components involved in the event: product family: dbs-linear leads, upn: m365db2202450, model: db-2202-45, serial: (b)(6), batch: (b)(6).Product family: dbs-linear leads, upn: m365db2202450, model: db-2202-45, serial: (b)(6), batch: (b)(6).
 
Event Description
It was reported that the deep brain stimulation (dbs) patient underwent a revision procedure to have the full dbs system explanted due to an infection.The patient was doing well postoperatively.Additional information was received that the infection originated in the suture behind the patients ear and spread to the entire dbs system.The patients suture would not dissolve and upon removing the suture, the patient expressed that the site became hot.Upon further investigation, the entire dbs system was compromised.The physician confirmed that the infection was caused by the suture and open wound at the suture site.The patient was placed on intravenous (iv) antibiotics for one month.A culture was performed and confirmed methicillin-resistant staphylococcus aureus (mssa).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERCISE GENUS
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key19029609
MDR Text Key339205975
Report Number3006630150-2024-02031
Device Sequence Number1
Product Code NHL
UDI-Device Identifier08714729985044
UDI-Public08714729985044
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date06/01/2024
Device Model NumberDB-1216
Device Catalogue NumberDB-1216
Device Lot Number543213
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/02/2024
Supplement Dates Manufacturer Received04/05/2024
Supplement Dates FDA Received04/30/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/02/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age42 YR
Patient SexMale
-
-