MEDOS INTERNATIONAL SàRL TRUESPAN 12 DEGREE PEEK; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 228151
Device Problems Device-Device Incompatibility (2919); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/22/2024
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H4: the device date of manufacture is unknown at this time.Udi: (b)(4).The photo investigation revealed that truespan 12 degree peek had the white sleeve removed from the shaft.The suture was still attached on the shaft.The implants weren¿t visible.No other anomalies could be observed.The overall complaint was unconfirmed as the observed condition of the truespan 12 degree peek would not contribute to the complained device issue.Based on the investigation findings, the potential cause for the reported condition could be traced to when not inserting the needle to the proper depth for deployment which could have caused blocking insertion, when this occurred resistance may have been felt and the trigger did not pull all the way.As per the instructions for use, it is necessary to use a calibrated probe, measure the width of the meniscal tissue to be repaired and set the adjustable depth.Also, it is important to fully squeeze the red deployment trigger while maintaining depth positioning to deliver the first implant.The implant is fully deployed when you hear an audible ¿click¿.Fully release the trigger after deployment, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes mitek quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Event Description
It was reported by the healthcare professional in colombia that during an unspecified surgical procedure the truespan 12 degree peek device does not come out and the gun is blocked as if the two peeks had been fired, the doctor takes out the gun with sutures and peek.He requests another suture and the procedure is completed without any inconvenience.There was no delay in the procedure reported.There were no adverse patient consequences reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUESPAN 12 DEGREE PEEK
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key19031296
MDR Text Key339236777
Report Number1221934-2024-01056
Device Sequence Number1
Product Code MBI
UDI-Device Identifier10886705026012
UDI-Public10886705026012
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K153667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number228151
Device Lot Number9L87816
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/26/2024
Initial Date FDA Received04/03/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-