Model Number 97810
Device Problems Migration or Expulsion of Device (1395); Charging Problem (2892); Communication or Transmission Problem (2896)
Patient Problems Pain (1994); Burning Sensation (2146)
Event Date 09/28/2022
Event Type  malfunction  
Manufacturer Narrative
G2: country: (b)(6).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was implan ted with an implantable neurostimulator (ins) for unknown indications for use.It was reported that patient had issues with recharging.Additional information was received.The manufacturer representative (rep)reported that the battery has moved.Patient can no longer charge the device and it will not register with the charging device.The ins failed to charge properly.Patient reported that after the first full charge back in (b)(6) 2022, it has trouble in charging properly.The recharger constantly gets disconnected and only charging up to 20%.Patient felt very frustrated because of this.Warm and painful whilst charging.New recharger issued on (b)(6) 2022.The problem resolved.Then on (b)(6) 2023, the ins was removed as patients wants a mitrofanoff.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key19033681
MDR Text Key339262192
Report Number3004209178-2024-08378
Device Sequence Number1
Product Code EZW
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/28/2022
Device Model Number97810
Device Catalogue Number97810
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/02/2024
Initial Date FDA Received04/03/2024
Date Device Manufactured06/08/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age21 YR
Patient SexFemale
Patient Weight82 KG
-
-