Model Number BI70000020
Device Problem Radiation Output Failure (4027)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/28/2024
Event Type  malfunction  
Manufacturer Narrative
No products have been returned to medtronic for analysis.Codes b17, c20, and d15 are applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
Medtronic received information regarding an imaging system being used during a sacroiliac and thoracolumbar procedure.It was reported that when an attempt was made to take an x-ray, there was no x-ray taken.The issue did not resolve after restarting the system, so the usage of the imaging system was discontinued.There was no reported delay to the procedure due to this issue.There was no reported impact on patient outcome.
 
Manufacturer Narrative
Continuation of d10: product id: bi71000447, lot number: unknown.H6: the system was serviced in the field and the hardware part was replaced.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Manufacturer Narrative
H3) component kit svc bi71000447 standalone and xrelay was returned for analysis.Functional testing determined that the component was malfunctioning causing the reported event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM IMAGING SYSTEM
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key19034262
MDR Text Key339272214
Report Number3004785967-2024-00224
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBI70000020
Device Catalogue NumberBI70000020
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/28/2024
Initial Date FDA Received04/03/2024
Supplement Dates Manufacturer Received03/30/2024
Supplement Dates FDA Received04/18/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/05/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
SEE H11...
-
-