• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FARAPULSE, INC. FARAWAVE PULSED FIELD ABLATION CATHETER, 35MM; CARDIAC IRREVERSIBLE ELECTROPORATION SYSTEM CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FARAPULSE, INC. FARAWAVE PULSED FIELD ABLATION CATHETER, 35MM; CARDIAC IRREVERSIBLE ELECTROPORATION SYSTEM CATHETER Back to Search Results
Lot Number 0008020937
Device Problem Improper Flow or Infusion (2954)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/28/2024
Event Type  malfunction  
Event Description
It was reported that a farawave ablation catheter 35mm presented a failure.During initial checks pre case it was running fine in flower.However during the procedure, the irrigation flow of the catheter stopped.No error messages at all just cessation of flow through the catheter.Liter bag was half full and pressure adequate, when the physician moved from collapsed to flower the flow stopped and returned once moving down to basket.The catheter was removed and device looked fine with nothing to indicate an issue.After venograms were performed the flow was checked to see if fluid is running in the sheath again, regularly checks were performed checking both lines are dripping suitably.The catheter had stopped running after the left sided veins when moved to flower shape.The catheter was changed no issue with next one under same lot number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FARAWAVE PULSED FIELD ABLATION CATHETER, 35MM
Type of Device
CARDIAC IRREVERSIBLE ELECTROPORATION SYSTEM CATHETER
Manufacturer (Section D)
FARAPULSE, INC.
3715 haven avenue
suite 110
menlo park CA 94025
Manufacturer (Section G)
AGILE MV INC
3700 st patrick st suite 102
montreal, qc H4E 1 A2
CA   H4E 1A2
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key19034275
MDR Text Key339268691
Report Number2124215-2024-19966
Device Sequence Number1
Product Code QZI
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 04/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Lot Number0008020937
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/02/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/11/2024
Initial Date FDA Received04/03/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/06/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-