Model Number 407658 |
Device Problems
Electrical /Electronic Property Problem (1198); Fracture (1260); High impedance (1291); Over-Sensing (1438); Capturing Problem (2891); Impedance Problem (2950)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/01/2024 |
Event Type
Injury
|
Event Description
|
It was reported that the right ventricular (rv) lead triggered a polarity switch and an alert for high unipolar pacing impedance was observed.High sensing integrity counter (sic) was also noted.A lead fracture was suspected.Monitored non-sustained ventricular tachycardia (vt) events showed non-physiologic intervals with what appeared to be oversensing.The device would be programed to the voo mode until the rv lead could be replaced.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
It was further reported that the rv lead also exhibited rising rv pacing impedance and rising rv thresholds.The rv lead was capped and replaced.
|
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
It was reported that the right ventricular (rv) lead triggered a polarity switch and an alert for high unipolar pacing impedance was observed.High sensing integrity counter (sic) was also noted.A lead fracture was suspected.Monitored non-sustained ventricular tachycardia (vt) events showed non-physiologic intervals with what appeared to be oversensing.The device would be programed to the voo mode until the rv lead could be replaced.No patient complications have been reported as a result of this event.(b)(6) 2024 it was further reported that the rv lead also exhibited rising rv pacing impedance and rising rv thresholds.The rv lead was capped and replaced.It was additionally reported that impedance varied between the normal baseline and a high undefined range.
|
|