Model Number 407658
Device Problems Electrical /Electronic Property Problem (1198); Fracture (1260); High impedance (1291); Over-Sensing (1438); Capturing Problem (2891); Impedance Problem (2950)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/01/2024
Event Type  Injury  
Event Description
It was reported that the right ventricular (rv) lead triggered a polarity switch and an alert for high unipolar pacing impedance was observed.High sensing integrity counter (sic) was also noted.A lead fracture was suspected.Monitored non-sustained ventricular tachycardia (vt) events showed non-physiologic intervals with what appeared to be oversensing.The device would be programed to the voo mode until the rv lead could be replaced.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was further reported that the rv lead also exhibited rising rv pacing impedance and rising rv thresholds.The rv lead was capped and replaced.
 
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was reported that the right ventricular (rv) lead triggered a polarity switch and an alert for high unipolar pacing impedance was observed.High sensing integrity counter (sic) was also noted.A lead fracture was suspected.Monitored non-sustained ventricular tachycardia (vt) events showed non-physiologic intervals with what appeared to be oversensing.The device would be programed to the voo mode until the rv lead could be replaced.No patient complications have been reported as a result of this event.(b)(6) 2024 it was further reported that the rv lead also exhibited rising rv pacing impedance and rising rv thresholds.The rv lead was capped and replaced.It was additionally reported that impedance varied between the normal baseline and a high undefined range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSUREFIX NOVUS MRI SURESCAN
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key19034501
MDR Text Key339271261
Report Number2649622-2024-09108
Device Sequence Number1
Product Code DTB
UDI-Device Identifier00885074200699
UDI-Public00885074200699
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P930039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/03/2017
Device Model Number407658
Device Catalogue Number407658
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/07/2024
Initial Date FDA Received04/03/2024
Supplement Dates Manufacturer Received04/04/2024
Supplement Dates FDA Received04/22/2024
Date Device Manufactured06/15/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
C6TR01 CRTP, 429688 LEAD
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age75 YR
Patient SexMale
-
-