DEPUY MITEK LLC US 4.5 HEALIX BR ANCHOR W/OCORD; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
Catalog Number 222229 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/24/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the product has not returned to depuy synthes mitek, however a photo was provided for review.The photo investigation revealed that 4.5 healix br anchor w/ocord was broken at the distal end of the anchor.No other anomalies could be observed.The overall complaint was confirmed as the observed condition of the 4.5 healix br anchor w/ocord would contribute to the complained device issue.Based on the investigation findings, the potential cause for the broken anchor can be traced to off axis insertion and levering during insertion.As per ifu- 109362 inserting the awl less than the specified depth, axial misalignment or levering with the anchor upon insertion, may result in anchor fracture, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes mitek quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H4: the date of manufacture was unknown.Udi: (b)(4).
|
|
Event Description
|
It was reported by the sales rep in china that during a rotator cuff repair surgical procedure on (b)(6) 2024 the 4.5 healix br anchor w/ocord device anchor was broken off.All broken parts were removed.Another like device was used to complete the procedure.There were no adverse patient consequences reported.No additional information was provided.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D4, h4: the expiration date and device manufacture date have been added.
|
|
Search Alerts/Recalls
|
|
|