• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C4063K
Device Problems Fluid/Blood Leak (1250); Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/08/2024
Event Type  malfunction  
Event Description
It was reported that the bladder of a large volume folfusor ruptured.The bladder rupture was described as, ¿the elastomer broke and the solution spilled inside the protective casing¿.This occurred while filling the device prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
E1: initial reporter first name:(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H4: the lot was manufactured july 18, 2023 - july 20, 2023.H10: the actual device was received for evaluation.Visual inspection was performed on the unit which observed liquid inside the housing indicating a leak has occurred.Further examination revealed the cause of leak inside the housing was due to missing a film-wrap located at the upper part of the bladder; no evidence of rupture was observed from the bladder.The film-wrap seals the bladder to the stress-member; therefore, when the film-wrap is missing, liquid would leak inside the housing.The reported leak inside the housing was verified.The cause of the missing film-wrap was related to manufacturing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key19035828
MDR Text Key339287299
Report Number1416980-2024-01498
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412080161
UDI-Public(01)00085412080161
Combination Product (y/n)N
Reporter Country CodeFI
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4063K
Device Lot Number23G008
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 03/08/2024
Initial Date FDA Received04/03/2024
Supplement Dates Manufacturer Received05/02/2024
Supplement Dates FDA Received05/03/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NA.
-
-