Model Number BF-1T150
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was observed during the device inspection, that the bronchovideoscope exhibited the coat of the connecting tube was peeled.There were no reports of patient involvement.
 
Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Corrected g2: health professional was inadvertently selected on the initial medwatch.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device was returned, and an evaluation was completed.During inspection, olympus found connecting tube has coating peeling (1mmsq or more).A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue were identified.Based on the results of the investigation, it is likely physical stress caused the coating to peel.However, a definitive root cause could not be determined.The following information is stated in the instructions for use (ifu): 3.2, "inspection of the endoscope"; 5, inspect the external surface of the entire insertion tube including the bending section and the distal end for dents, bulges, swelling, peeling, scratching, holes, sagging, transformation, bends, adhesion of foreign body, dropout of parts, any protruding objects or other irregularities".Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRONCHOVIDEOSCOPE
Type of Device
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19037798
MDR Text Key339333349
Report Number9610595-2024-06987
Device Sequence Number1
Product Code EOQ
UDI-Device Identifier04953170288968
UDI-Public04953170288968
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K023984
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberBF-1T150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/15/2024
Initial Date FDA Received04/03/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/05/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse