Device Problem
Off-Label Use (1494)
|
Patient Problem
Syncope/Fainting (4411)
|
Event Date 03/18/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
|
|
Event Description
|
It was reported that during an irreversible electroporation using a farawave catheter the patient experienced a vasovagal response.During the procedure a posterior wall ablation was performed, posterior wall ablations are considered off-label use for the farawave catheter.The farawave's instructions for use (ifu) sates that "the farawave catheter is indicated for the isolation of pulmonary veins in the treatment of paroxysmal atrial fibrillation".After the posterior wall ablation validation mapping found a delayed vagal response, as the patient was "unable to 1:1 conduct from atria to ventricles".The patient was able to "conduct down to 360ms" after an isuprel infusion.It is not currently known if the device will be returned for analysis.
|
|
Event Description
|
It was reported that during an irreversible electroporation using a farawave catheter the patient experienced a vasovagal response.During the procedure a posterior wall ablation was performed, posterior wall ablations are considered off-label use for the farawave catheter.The farawave's instructions for use (ifu) sates that "the farawave catheter is indicated for the isolation of pulmonary veins in the treatment of paroxysmal atrial fibrillation".After the posterior wall ablation validation mapping found a delayed vagal response, as the patient was "unable to 1:1 conduct from atria to ventricles".The patient was able to "conduct down to 360ms" after an isuprel infusion.It is not currently known if the device will be returned for analysis.It was further reported that the device was disposed of following the procedure and that the patient recovered without the need for any further treatments or interventions.
|
|
Manufacturer Narrative
|
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
|
|
Search Alerts/Recalls
|