Model Number 0672 |
Device Problems
Failure to Capture (1081); Fracture (1260); Mechanical Problem (1384); Contamination /Decontamination Problem (2895); Device Dislodged or Dislocated (2923); Positioning Problem (3009); High Capture Threshold (3266)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/14/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The product has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental report will be filed.
|
|
Event Description
|
It was reported that this right ventricular (rv) lead was not successfully implanted due to helix stopped moving during procedure.Also, this lead was repositioned several times due to lead exhibited high threshold.Also, premature ventricular contraction (pvc) suddenly appeared frequently, and lead was expected to dislodge and confirmed it with fluoroscopy.The ventricle was measured using pacing system analyzer (psa) but it was not detected.In addition, when the helix was checked, a piece of tissue was recognized.Furthermore, physician opinion was that the lead was fractured due to the repeated screwing in and out to position the lead.A new lead was implanted.No adverse patient effects were reported.
|
|
Event Description
|
It was reported that during an implant procedure, this right ventricular (rv) lead was attempted to be implanted, however, high capture thresholds were observed, therefore, it was repositioned several times.After multiple placements, the threshold value stabilized, however, this lead dislodged confirming this by fluoroscopy and loss of capture (loc) was observed.Repositioning was again attempted; however, helix issues were exhibited.The physician opted to remove this lead.Visual inspection noted tissue within the helix.The lead was washed, and the tissue removed, however, the helix was not able to be retracted, therefore, a new lead was implanted.A lead fracture was suspected by the physician.No adverse patient effects were reported.This lead has been received for analysis.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Testing was completed to assess lead electrical performance and inner and outer insulation integrity.Measurements throughout these tests were within normal limits.Visual inspection of the lead found that the helix mechanism was retracted and dried blood/tissue was noted in the helix housing.Visual inspection also showed that the insulation was bunched, and conductor coils were deformed approx., 170-255mm from the terminal end.This is consistent with the lead being placed through a constricted area.This type of damage has been deemed a design issue.
|
|
Event Description
|
It was reported that during an implant procedure, this right ventricular (rv) lead was attempted to be implanted, however, high capture thresholds were observed, therefore, it was repositioned several times.After multiple placements, the threshold value stabilized, however, this lead dislodged confirming this by fluoroscopy and loss of capture (loc) was observed.Repositioning was again attempted; however, helix issues were exhibited.The physician opted to remove this lead.Visual inspection noted tissue within the helix.The lead was washed, and the tissue removed, however, the helix was not able to be retracted.A new lead was implanted.A lead fracture was suspected by the physician.No adverse patient effects were reported.This lead has been received for analysis.
|
|
Search Alerts/Recalls
|
|