Catalog Number 07671687016
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/22/2024
Event Type  malfunction  
Event Description
We received an allegation of questionable inr results for one patient tested with the coaguchek inrange meter compared to the laboratory method.On (b)(6) 2024: the laboratory result from the stago neoptimal laboratory method was 2.4 inr.  the meter result from the reporter's coaguchek inrange meter was 3.4 inr.The time between measurements was <3 hours.On (b)(6) 2024: the laboratory result from an unknown laboratory method was 2.5 inr.  the meter result from the reporter's coaguchek inrange meter was 3.2 inr.The meter result from an unspecified coaguchek retention meter was 3.5 inr.The time between measurements was not provided.The patient's therapeutic range is 2 to 3 inr.   .
 
Manufacturer Narrative
The serial number of the customer's coaguchek inrange meter is (b)(6).The serial number of the coaguchek retention meter was not provided.The patient was taking antibiotics at the time of the report.Product labeling states "the action of oral anticoagulants (coumarin derivatives) can be increased or weakened when other medication is taken simultaneously (e.G.Antibiotics, but also prescription-free medication like pain relievers, antirheumatic medication and medication against influenza).This, in turn, can also lead to either an increase or a decrease in prothrombin time (inr).If other medication is taken, it is recommended that the prothrombin time be checked more frequently and that the anticoagulant dose be subsequently adjusted." on a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Per product labeling, "coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods." e3 - occupation is patient/consumer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS PT TEST PST
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key19040808
MDR Text Key340124877
Report Number1823260-2024-01030
Device Sequence Number1
Product Code GJS
Combination Product (y/n)Y
Reporter Country CodeFR
PMA/PMN Number
K170960
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number07671687016
Device Lot Number67810016
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/11/2024
Initial Date FDA Received04/04/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ANTIBIOTICS; FLUINDION
-
-