The serial number of the customer's coaguchek inrange meter is (b)(6).The serial number of the coaguchek retention meter was not provided.The patient was taking antibiotics at the time of the report.Product labeling states "the action of oral anticoagulants (coumarin derivatives) can be increased or weakened when other medication is taken simultaneously (e.G.Antibiotics, but also prescription-free medication like pain relievers, antirheumatic medication and medication against influenza).This, in turn, can also lead to either an increase or a decrease in prothrombin time (inr).If other medication is taken, it is recommended that the prothrombin time be checked more frequently and that the anticoagulant dose be subsequently adjusted." on a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Per product labeling, "coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods." e3 - occupation is patient/consumer.
|