Model Number STRIP, CVS 100CT KETONE
Device Problem Material Discolored (1170)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2024
Event Type  malfunction  
Manufacturer Narrative
Internal report reference number: (b)(4).Ketone test strips were returned.Defect found on returned strips: physical defect of strips; discolored grey pads.Retention testing was performed using test strips from the same lot.Retention strip lot tested within specifications.Root cause: rc-072: vial left open for an extended period of time.Note: manufacturer contacted customer in a follow-up call on 19-mar-2024 to ensure the replacement products resolved the initial concern - able to establish contact with customer who stated replacement products resolved initial concern.
 
Event Description
Consumer reported complaint for the trueplus ketone test strips.Customer stated that when he opened the bottles of ketone strips, the padding was gray.Customer advised that he just purchased the products today, (b)(6) 2024.Customer did not attempt to use any of the strips from the vial.The customer feels well and did not report any symptoms and medical attention related to the use of the product was not reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KETONE
Type of Device
NITROPRUSSIDE, KETONES (URINARY, NON-QUANT.)
Manufacturer (Section D)
TRIVIDIA HEALTH INC
2400 nw 55th court
fort lauderdale FL 33309
Manufacturer Contact
karen devincent
2400 nw 55th court
fort lauderdale, FL 33309
MDR Report Key19043024
MDR Text Key340123847
Report Number1000113657-2024-00135
Device Sequence Number1
Product Code JIN
UDI-Device Identifier021292008178
UDI-Public(01)021292008178
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberSTRIP, CVS 100CT KETONE
Device Lot NumberAA673
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2024
Distributor Facility Aware Date03/11/2024
Initial Date Manufacturer Received 03/11/2024
Initial Date FDA Received04/04/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-