Model Number STRIP, CVS 100CT KETONE |
Device Problem
Material Discolored (1170)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/11/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal report reference number: (b)(4).Ketone test strips were returned.Defect found on returned strips: physical defect of strips; discolored grey pads.Retention testing was performed using test strips from the same lot.Retention strip lot tested within specifications.Root cause: rc-072: vial left open for an extended period of time.Note: manufacturer contacted customer in a follow-up call on 19-mar-2024 to ensure the replacement products resolved the initial concern - able to establish contact with customer who stated replacement products resolved initial concern.
|
|
Event Description
|
Consumer reported complaint for the trueplus ketone test strips.Customer stated that when he opened the bottles of ketone strips, the padding was gray.Customer advised that he just purchased the products today, (b)(6) 2024.Customer did not attempt to use any of the strips from the vial.The customer feels well and did not report any symptoms and medical attention related to the use of the product was not reported.
|
|