There was an allegation of questionable elecsys pth immunoassay and elecsys total psa results for 2 patient samples on a cobas e 411 immunoassay analyzer.The customer was prompted to repeat the samples as their qc was out of range and after repeating the qc and obtaining acceptable results, they repeated previously run patient samples.For sample 1, the initial psa result was 0.006 ng/ml with flag.The sample was repeated and the repeat result was 0.269 ng/ml.For sample 2, the initial pth result was 2.40 pg/ml with flag.The sample was repeated and the repeat result was 43.87 pg/ml.
|