It was reported that a mitraclip procedure was performed to treat degenerative mitral regurgitation (mr) with a grade of 4+, a mean pressure gradient of 1.1mmhg, several subvalvular chordae, large p2 into prolapsed posterior leaflet, small p2 flail, rotated heart, esophagus off midline, and low transseptal height for procedure.A mitraclip xtw was implanted.Another mitraclip xtw (30803a1079) was inserted and was advanced to the mitral valve.However, during deployment, when the grippers were pulled back, the clip spun and detached from the posterior mitral leaflet (pml), causing damage to the pml.The clip was implanted.Pre deployment images were reviewed, and the clip was observed to be properly aligned and leaflet insertion was confirmed.It was likely that a chord had also been grasped, which was not appreciated, causing the clip to detach from the pml.To stabilize the detached clip and further reduce the mr, another mitraclip xtw (30803a1081) was inserted, and grasping was performed.A successful grasp was achieved, but the gradient increased between 8mmhg and 10mmhg.After ungrasping the leaflets, the pressure gradient returned to baseline.The clip was removed from the patient.There were no reported patient adverse effects related to the increase in pressure gradient.It was noted that difficulties visualizing the second and third clip during the procedure occurred.The procedure was completed with two clips implanted, reducing the mr to a grade of 3-4+.The final mean pressure gradient was at 4.5mmhg.There was no clinically significant delay in the procedure.On (b)(6), 2024, the patient underwent mitral valve replacement surgery.
|
In this case, two clips only (without the clip delivery system) were returned.The gripper being misaligned (product quality problem) was also confirmed.The reported entrapment of device- clip caught on chordae, image resolution poor n/a and incomplete coaptation-intraprocedure during use could not be replicated in a testing environment.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no similar complaints from the lot.All information was investigated and based on the information provided and the analysis of the returned clip, the reported product quality problem associated with misaligned gripper appears to be related to the gripper being caught on the clip cover and giving it the appearance to be misaligned.A cause for how it became caught however, cannot be determined.Additionally, a cause for the reported slda cannot be determined as it cannot be definitively confirmed whether the gripper being caught on the clip cover or the clip becoming entangled with the anatomy contributed to the slda.Image resolution poor is related to procedural conditions associated with difficulties visualizing the clip during the procedure.The reported entrapment of device associated with a chord likely being grasped during procedure appears to be related to patient conditions (several subvalvular chordae, rotated heart and low transseptal height).Unspecified tissue injury (damage to the posterior mitral leaflet (pml)) appears to be due to the slda.Tissue damage is listed as a known possible complication associated with mitraclip procedures.Unexpected medical intervention, surgical intervention and hospitalization were a result of case-specific circumstances.There is no indication of a product issue with respect to manufacture, design or labeling.The additional cds0706-xtw device referenced in b5 is filed under separate medwatch report number.
|