The cobas e411 disk serial number was (b)(6).The customer stated that qc was within the assigned ranges on the day of the event.The field service engineer observed that the tip of the reagent/sample pipetter was slightly dirty and 2 pinch tubes needed replacement.He cleaned the pipetter and replaced the 2 pinch tubes.He also performed measuring cell preparation maintenance.Assay performance check, calibration, and qc were performed and they were acceptable.Comparison with already processed samples was performed and the results were acceptable.The investigation is ongoing.
|
Sections d1, df2a, d2b, d4, g1 and g4 were updated.A general reagent problem can be excluded because the qc prior to the event was within ranges.The field service engineer (fse) adjusted the liquid level detection voltage, the sample/reagent pipettor, and the sipper.The fse replaced the reading cell and the pipes related to the reading circuit.He also performed a system volume check and voltage adjustment.The service actions (adjusting the liquid level detection voltage, the sample/reagent pipettor, the sipper, replacing the reading cell and the pipes related to the reading circuit, and performing a system volume check and voltage adjustment) resolved the issue.No further issues were reported afterward.The cause of the event could not be determined.
|