• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number PLC271000J
Device Problems Patient-Device Incompatibility (2682); Device-Device Incompatibility (2919)
Event Date 03/15/2024
Event Type  Injury  
Event Description
On (b)(6) 2020, this patient underwent an endovascular treatment of an abdominal aortic aneurysm using gore® excluder® aaa endoprosthesis and gore® excluder® iliac branch endoprosthesis.On an unknown date, a type iii endoleak from the junction between the iliac branch component and the contralateral leg component was observed.Reportedly, the contralateral leg component was completely disengaged from the iliac branch component.On (b)(6) 2024, a reintervention was performed, two additional stent grafts were placed, the endoleak was resolved.The patient tolerated the procedure.Reportedly, it seems the proximal edge of the internal iliac component was implanted about 1cm beyond the iliac branch component flow divider, which shortened the overlap length and caused the contralateral leg component to be disengaged.
 
Manufacturer Narrative
H3: code "other" was selected as the medical device remains implanted.Return not possible.H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.According to the gore® excluder® aaa endoprosthesis instructions for use(ifu), adverse events that may occur and / or require intervention or additional intraoperative procedure time include, but are not limited to: endoleak.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Manufacturer Narrative
Emdr section h6, medical device problem code updated to a1702.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Type of Device
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL SILICON VALLEY B/P
2890 de la cruz blvd.
santa clara CA 95050
Manufacturer Contact
hiroko sakamiya
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key19048625
MDR Text Key339431448
Report Number3013164176-2024-02033
Device Sequence Number1
Product Code MIH
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P020004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/14/2020
Device Catalogue NumberPLC271000J
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/15/2024
Initial Date FDA Received04/05/2024
Supplement Dates Manufacturer Received03/15/2024
Supplement Dates FDA Received04/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/15/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age80 YR
Patient SexMale
-
-