DATEX-OHMEDA, INC. AISYS CS2; ANESTHESIA GAS MACHINE Back to Search Results
Model Number 1011-9050-000
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/07/2024
Event Type  malfunction  
Event Description
It was reported that there was a malfunction resulting in system shutdown during a case.The patient was switched to manual ventilation.There was no patient injury.
 
Manufacturer Narrative
Ge healthcareâs investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.Block a: there was no patient information available at time of mdr submission.Legal manufacturer: hcs madison 3030 ohmeda dr, usa, madison, wi.53718.
 
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The display was replaced to resolve the issue.Block a: no patient information provided to date after calls to customer 04/24/24 and 05/06/24.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AISYS CS2
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
anthony amenson
3030 ohmeda drive
madison, WI 53718
MDR Report Key19050999
MDR Text Key339572869
Report Number2112667-2024-01853
Device Sequence Number1
Product Code BSZ
UDI-Device Identifier00840682102292
UDI-Public0100840682102292
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 05/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1011-9050-000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/08/2024
Initial Date FDA Received04/05/2024
Supplement Dates Manufacturer Received04/21/2024
Supplement Dates FDA Received05/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/17/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-