• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS LUCERA ELITE GASTROINTESTINAL VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS LUCERA ELITE GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-HQ290
Device Problem Device Reprocessing Problem (1091)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the endoscope reprocessor water filters were replaced once every 2 years instead of once every 2 months, and the user facility staff irregularly disinfected the water supply pipes approximately once every 1-2 months after replacing the water filters.As a result, the scope was insufficiently reprocessed.No patient infections or injuries reported.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 4 years since the subject device was manufactured.Based on the results of the investigation it is likely that the reprocessing for the endoscope was insufficient since the filter was not replaced.However, the root cause was unable to be specified.The event can be detected/prevented by following the instructions for use: 7.3 replacing the water filter (maj-2317) replace the water filter periodically, at least once in two months to prevent contamination of the rinse water.[warning] ¿ after replacing the water filter, be sure to perform the operation described in section 7.4, ¿disinfecting the water supply piping¿ to prevent multiplication of miscellaneous germs and staining inside the water supply pipes.Failure to perform this operation could result in contamination of the equipment piping and/or the scope, preventing effective reprocessing.¿replace the water filter at least once in two months.Otherwise, miscellaneous germs and stains in the water may contaminate the equipment and/or the endoscopes and prevent effective reprocessing.7.4 disinfecting the water supply piping.Disinfection of water supply piping is required in the following cases: ¿ before using this equipment for the first time (after installation of the water filter set).¿ immediately after replacement of the water filter.¿ whenever bacteria are identified in the water supply piping.¿ before using the equipment when it has not been used for more than 14 days.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA ELITE GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19051442
MDR Text Key339776370
Report Number9610595-2024-07130
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIF-HQ290
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/01/2024
Initial Date FDA Received04/05/2024
Supplement Dates Manufacturer Received04/25/2024
Supplement Dates FDA Received04/30/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/12/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
OER-6-OLYMPUS ENDOSCOPE REPROCESSORSN. (B)(6)
-
-