BAXTER HEALTHCARE CORPORATION PRISMAFLEX SETS (ST); DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 955596
Device Problems Fluid/Blood Leak (1250); Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/08/2024
Event Type  malfunction  
Event Description
It was reported that a prismaflex st150 set leaked at the connection of the effluent line to the support plate.Closer inspection revealed that the connection at the interface was not tight.This occurred during priming for hemodialysis treatment.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Prismaflex st150 set c has been temporarily approved for use in the us under emergency use authorization eua200704 to deliver crrt to treat patients in an acute care environment during the covid-19 pandemic.The actual device was not available; however, two (2) photographs of the sample were provided for evaluation.The photographs were reviewed and showed the effluent line pre pump is cut inside the support plate.The reported condition was verified.The most probable cause was due to a shock and an exposition to low temperature that occurred during transport by local carrier.A nonconformance has been opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX SETS (ST)
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key19052053
MDR Text Key339497800
Report Number8010182-2024-00149
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414124403
UDI-Public(01)07332414124403
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 04/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number955596
Device Lot Number23G0050CB
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/21/2024
Initial Date FDA Received04/05/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-