MAUDE Adverse Event Report: GYRUS ACMI, INC. SHOCKPULSE-SE LITHOTRIPSY SYSTEM; LITHOTRIPTOR, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC. SHOCKPULSE-SE LITHOTRIPSY SYSTEM; LITHOTRIPTOR, ULTRASONIC Back to Search Results
Model Number SPL-S
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/07/2024
Event Type  malfunction  
Manufacturer Narrative
Three attempts were performed to obtain additional information, but no response was received from the customer.The device was returned to olympus for evaluation and the customer¿s allegation was confirmed.The device evaluation found the connector pins were broken and the transducer cannot be plugged in.Based on the results of the investigation and information obtained from this complaint, the cause was traced to a component failure.A device history review revealed findings/no issues that could have caused or contributed to the reported issue.Olympus will continue to monitor the field performance of this device.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the subject device "appears to have broken connector pins.Would not work".The issue happened during an unspecified therapeutic procedure.There was no report of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHOCKPULSE-SE LITHOTRIPSY SYSTEM
Type of Device
LITHOTRIPTOR, ULTRASONIC
Manufacturer (Section D)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19052058
MDR Text Key339497864
Report Number3011050570-2024-10057
Device Sequence Number1
Product Code FEO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142428
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSPL-S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/07/2024
Initial Date FDA Received04/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/28/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-