Catalog Number UNK_WWA
Device Problems Osseointegration Problem (3003); Migration (4003)
Patient Problems Cyst(s) (1800); Inadequate Osseointegration (2646); Implant Pain (4561)
Event Date 01/19/2024
Event Type  Injury  
Manufacturer Narrative
The reported event could be confirmed, based on available information, ct scans and medical expert opinion.The device inspection was not possible as the product was not returned for investigation.The device history record could not be reviewed because the affected device was not returned, and the lot number was not communicated.A review of the labeling did not indicate any abnormalities.Upon further investigation of the ct scans by healthcare professionals the following was observed tibial (construct): the tibial tray shows significant radiolucence and some small cysts.Loosening is likely and a little bit of subsidence seems to have occurred.Pe: the pe is not separated from the tibial tray.There is no loosening or breakage visible.Talar (construct): the talar component shows clear radiolucence and very large instable anterior cysts.The component is loosened and anteriorly migrated/subsided.Based on investigation, the root cause was attributed to a patient related issue.The failure was identified by radiolucency and cysts which contributed to subsidence of the tibial and talar construct.If device is returned or any further information is provided, the investigation report will be reassessed.H3 other text : device remains implanted in patient.
 
Event Description
The digital stryker prophecy team received a ct scan indicating that the patient may require a revision surgery due to pain and subsidence of the talar component.The x-ray and ct scan is showing significant cyst formation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN TIBIAL COMPONENT
Type of Device
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key19055472
MDR Text Key339550839
Report Number3010667733-2024-00187
Device Sequence Number1
Product Code HSN
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 04/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Catalogue NumberUNK_WWA
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/11/2024
Initial Date FDA Received04/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexMale
-
-