Model Number 30MM CENTRAL SCREW, MODULAR
Device Problem Positioning Failure (1158)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/24/2023
Event Type  malfunction  
Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Complaint is confirmed.Visual evaluation, the edges of the hex screw and base of the central screw had damage.Functional testing identified that the returned device did not work as required.The probable cause of this condition of damage can be that the hex tip must be aligned between the components in order for the taper to engage and rotate the components until the hex features align, resulting in a tactile coupling and were not engaged properly.The most likely cause is attributed to misuse due to user error.
 
Event Description
It was reported that during an univers revers mgs surgery screw which is pressed into the plate did not hold.There was no harm for patient, operator or third party reported.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
30MM CENTRAL SCREW, MODULAR
Type of Device
SHOULDR PROSTH, REVERSE CONFIG
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19055896
MDR Text Key339635880
Report Number1220246-2024-01851
Device Sequence Number1
Product Code PHX
UDI-Device Identifier00888867296572
UDI-Public00888867296572
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K173900
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 04/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number30MM CENTRAL SCREW, MODULAR
Device Catalogue NumberAR-9561-30S
Device Lot Number14975759
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2023
Initial Date Manufacturer Received 01/25/2023
Initial Date FDA Received04/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/28/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-